Occlusion, air-in-line, door open, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm.
Real-time infused volume, bolus rate, bolus volume, KVO rate, automatic power switching, mute key, purge, bolus, system memory, key locker, and flow rate adjustment without stopping.
| Model | SY-G076-2 |
| Pumping Mechanism | Curvilinear peristaltic |
| Flow Rate | 0.1-1500 ml/h (in 0.1 ml/h increments) |
| Purge, Bolus | 100-1500 ml/h (in 1 ml/h increments) |
| Accuracy | ±3% |
| VTBI | 1-9999 ml |
| KVO Rate | 0.1-5 ml/h |
| Occlusion Sensitivity | High, medium, low |
| Air-in-line Detection | Ultrasonic detector |
| Battery Life | 5 hours at 30 ml/h |
| Weight | 2.5 kg |
| Safety Classification | Class 1, type CF |
| Size | 174*126*215 mm |
The device supports multiple modes including ml/h, drop/min, and time-based infusion to meet different clinical requirements.
Yes, the SY-G076-2 is designed to be compatible with IV sets of any standard, providing flexibility in medical supplies.
The rechargeable battery provides up to 5 hours of continuous operation at a flow rate of 30 ml/h.
It features a 10-way alarm system, including air-in-line detection via an ultrasonic detector, occlusion sensitivity, and system malfunction alerts.
Yes, one of its advanced features allows users to change the flow rate without stopping the pump, ensuring continuous treatment.
The flow rate is highly adjustable, ranging from 0.1 ml/h up to 1500 ml/h, with precise 0.1 ml/h increments.